Effluent quality: resistivity 18.2 MΩ·cm @25 °C, TOC ≤ 5 ppb or 1–3 ppb.

 

A. Complete Process Flow (Where UV Fits In)

A typical ultrapure water system is a multi-stage train of “tap water → purified water → Type 1 ultrapure water”. UV photo-oxidation sits in the final polishing stage, not in the front-end disinfection position.

1. Pre-treatment: pre-filter / activated carbon / softening — removes residual chlorine, large particles and hardness.

2. RO (reverse osmosis): removes salts, colloids, most organics and microorganisms.

3. EDI (electrodeionization): continuous demineralization, producing RO / purified-water grade water.

4. Final stage:

l Polishing cartridge 1: resin, removes residual ions;

l Dual-wavelength UV photo-oxidation unit: degrades trace organics and lowers TOC;

l Polishing cartridge 2: activated carbon, adsorbs organics, particles and pyrogens;

l Ultrafiltration (UF): removes endotoxins, nucleases and particles (optional configuration).

5. Circulation loop back to the tank → point of use.

Positioning: UV does not carry the main demineralization duty — RO / EDI / resin have already removed most salts and organics. UV only photo-oxidizes the remaining trace neutral organics into charged ions, which the downstream resin then removes, finally pushing TOC below 5 ppb.

 

B. Photochemical Mechanism of UV TOC Removal (Dual-Wavelength Synergy)

It uses a low-pressure mercury dual-wavelength lamp: 185 nm (vacuum UV, VUV) + 254 nm (germicidal UV), degrading organics through two pathways.

Pathway 1: Direct Bond Scission at 185 nm

Photons at 185 nm carry high energy (about 6.7 eV) and directly break the C–C and C–H bonds of organics, cleaving large molecules into small fragments.

Pathway 2: Combined 185 + 254 nm Generation of Hydroxyl Radicals (Main Oxidation Pathway)

1. Water photolysis (185 nm): H₂O ──185nm──→ OH· + H·

2. Ozone formation from dissolved oxygen (185 nm): ³⁄₂ O₂ ──185nm──→ O₃

3. Ozone decomposition (254 nm): O₃ ──254nm──→ O₂ + O·

4. Reactive oxygen species regenerate hydroxyl radicals: O· + H₂O ──254nm──→ 2 OH· + O₂

5. OH· attacks organics with strong oxidizing power and finally mineralizes them into inorganic species: organics ──OH·──→ CO₂ + H₂O. The CO₂ dissolves in water to form carbonate / bicarbonate (charged ions), which are adsorbed and removed by the downstream ion-exchange resin.

In one sentence: 185 nm does the “cutting” and generates the reactive species, while 254 nm sustains the oxidation chain and disinfects — neither can be omitted. 254 nm alone kills bacteria but cannot push TOC down.

特殊灯管 (1).jpg 特殊灯管 (4).jpg

 

C. UV Chamber Structure (How Water Flow and the Lamp Work Together)

• The lamp is enclosed in a high-purity UV-transparent quartz sleeve. Purified water flows through the annular path between the sleeve and the housing and is irradiated 360° around its circumference.

• 185 nm belongs to the vacuum UV range and is absorbed by ordinary quartz; TOC lamp sleeves must therefore use a special synthetic silica material that is highly transparent at 185 nm (does not block 185 nm). This is the key material difference from an ordinary 254 nm germicidal lamp.

• Dissolved oxygen in pure water is low, which limits radical yield, so this unit is normally installed in the circulation loop for repeated irradiation rather than a single pass.

 

D. Principle of In-Line TOC Measurement (Dual Use of the Same Lamp)

1. After each water draw, the product water bypasses the final polishing cartridge and flows directly through the UV lamp via the circulation loop.

2. The lamp oxidizes neutral organics into charged molecules, so the conductivity of the water rises accordingly.

3. A built-in resistivity / conductivity sensor detects this change, and an algorithm converts it into a TOC value.

4. After water draw, the touch screen displays TOC: XXX ppb.

 

E. Key Technical Parameters and Consumables

Light source type

Low-pressure mercury dual-wavelength photo-oxidation lamp (Photooxidation UV Lamp)

Main wavelengths

185 nm + 254 nm (185 nm degrades organics, 254 nm disinfects and continues oxidation)

Power

Selected according to the flow rate of the laboratory ultrapure water machine — LIGHTBEST DURAL UV lamps

Target TOC

System outlet < 5 ppb after oxidation

Lamp replacement interval

4500 H (actual service life is far longer, but effectiveness may decline beyond this point)

Sleeve

Special synthetic silica material (highly transparent at 185 nm)

特殊灯管 (3).jpg 特殊灯管 (2).jpg

 

F. Division of Labour of the Two UV Lamps in the System

Item

Germicidal UV lamp

Organic-decomposition (TOC) UV lamp

Wavelength

254 nm

185 nm + 254 nm dual wavelength

Function

Damages microbial DNA, inhibits bacteria / algae growth

Photo-oxidizes trace organics, lowers TOC

Installation position

Tank / circulation loop

Final polishing stage

Quartz requirement

Transparent at 254 nm is sufficient

Must be highly transparent at 185 nm vacuum UV

Replacement part

Disinfection consumable

Photo-oxidation consumable, directly tied to the TOC result

 

 

G. Key Points About the Lamp Itself

• 185 nm transparency is the core barrier of a TOC lamp: whether 185 nm is effectively transmitted, whether the electrode emission material is stable, and whether the lamp’s useful life reaches 4500 H — these are the decisive points for this type of instrument spare part. Ordinary water-treatment and ozone-generating UV lamps look much the same from the outside, but because of material and process limitations their 185 nm transmittance is low and drops sharply after 500–1000 h, seriously degrading TOC oxidation performance. Screening and selection must be done with great care.

• Instrument-matched lamp ≠ industrial water-treatment lamp: different power ratings, dual wavelength, low noise, long life, and extremely high requirements on batch-to-batch consistency; they are normally supplied as a complete set with lead wires / connector / quartz sleeve.

• Effluent TOC is a system-level result: it is not determined by a single lamp — the resin, the circulation loop and the feed-water TOC all affect the final reading. In external benchmarking, state clearly that “the lamp is the photo-oxidation step, not a stand-alone TOC removal device”.